Description
Features:
- With embedded computer and optimized optical system ,convenient and easy to use.
- Operate with both 4.3” touch panel screen and keypads.
- Abundant but simple instructions lead you no manual operation;
- Basic photometric mode,Quantitative analysis,Kinetics test,Wavelength scan,DNA/Protein and multi-wavelengths test are available;
- Alternative fitting and correction methods for Quantitative analysis;
- Optional built in thermal printer output data and curve;
- Up to 1000 test data can be saved in local memory and transfer to USB storage;
- Optional ADX’s Windows®-based PC application software provides more features.Kinetics test(Enzyme reaction rate calculation),DNA/Protein test and Wavelength scan are included.
Specifications:
Models | Alpha-1506 Plus |
Wavelength Range | 190 – 1100nm |
Spectral Bandwidth | 2nm/4nm(optional) |
Optical System | Single beam, Littrow,Grating 1200 line/mm |
Wavelength Accuracy | ±0.5nm |
Wavelength Repeatability | ±0.2nm |
Wavelength Resolution | 0.1nm |
Photometric Range | 0.3 to 3A ; 0-200%T; -9999 to 9999C |
Photometric Accuracy | ±0.004A@0.5A , ±0.5%T(0~100%T) |
Photometric Repeatability | ±0.002A@0.5A , ±0.3%T(0~100%T) |
Stray Light | ≤0.05%T@220nm&340nm |
Baseline flatness | ±0.002A(PC running software required) |
Stability | ±0.002A/hr @500nm ,0A |
Data Output | 480×272 touch panel LCD |
built-in printer & USB storage(optional) | |
Slave USB Connecting to computer | |
Size | 570×460×230 mm |
Weight | 14.5kg |
Reviews